Krantz Components

Filter systemin the studio

SCF hightec

Modular HEPA filter system

SCF HIGHTEC Modular HEPA filter system by Krantz
Image: Krantz Components
SCF hightecFilter system · high-containment
Leak-tightness class
housing and filter seat to DIN 25 496 table 3; test groove per elementp. 3
Application
laboratories, chemical, pharmaceutical and nuclear industryp. 2
Volume flow
up to 48,000 m³/h nominal (higher on request)
Source · E1.1.2Krantz, HEPA filter system SCFhightec, E1.1.2 07.2013

How the SCF hightec works.

Applications: modular, space-saving safe-change filter housing, with the housing arranged vertically or horizontally (krantz.de).

Krantz, HEPA filter system SCFhightec, E1.1.2 07.2013.

Technical data SCF hightec

Volume flow
up to 48,000 m³/h nominal

Krantz, HEPA filter system SCFhightec, E1.1.2 07.2013.

Page numbers are PDF pages of the data sheet.

Download documents

Request CAD/BIM

We will send you the available CAD or BIM files for the SCF hightec, or call you to hear exactly what you need.

We use your details only to handle this request and do not store them on the website. More in our privacy statement.

Air

Specifications of the SCF hightec with the source of each value · Air
PropertyValue · Source
Volume flow up to 48,000 m³/h nominal (higher on request)
Pressure difference ± 6,000 Pa permissible
Temperature 90 °C

Design

Specifications of the SCF hightec with the source of each value · Design
PropertyValue · Source
Construction compact, modular kit; operation from one side; vertical or horizontal
Leak-tightness housing and filter seat to DIN 25 496 (leakage < 3·10⁻⁵ of nominal volume flow at 2,000 Pa)
Filter class fijnstof F (EN 779, 610 x 610 x 78 mm) en HEPA H (EN 1822, 610 x 610 x 292 mm), bijv. H13 99.95% p. 8
Leak-tightness class housing and filter seat to DIN 25 496 table 3; test groove per element p. 3
Application laboratories, chemical, pharmaceutical and nuclear industry p. 2

Your host

Tell Pia about your space.

Pia is Peek's digital host. Describe your hall or building, and she will show you the Krantz components that fit.

I'm Pia. Shall I show you around?

Pia is an AI avatar, not a person.

with sound and microphone

Your voice and the conversation go via Runway in the United States, where privacy protection is weaker than in the EU; we do not yet have a data processing agreement with them. By clicking the start button you give your explicit consent to this (see the privacy statement).

Not live. Pia only starts when you click.

Prefer a call? Leave your name and number and we'll call you back. Or chat with Pia using the button at the bottom right.

Or call us: +31 (0)73 820 02 88

Video call with Pia

What do you think of the new site?

Your answer helps us improve the site.

Your rating

1 = needs a lot of work, 5 = very good

We receive your rating, your text, the page you were on and, only if you fill it in, your email address, by email at Peek. More in the privacy statement.